.JFIF
1/1
Protocol: Temporary Pacemaker Verification and Troubleshooting
figure
posted on 2022-10-06, 18:44 authored by Jill Ley(Figure 1)
(Figure 2)
Determining the Stimulation Threshold
- NOTE: a paced rhythm must be present and patient stable
- From 100% pacing, gradually decrease output for A or V, while watching monitor until you lose capture
- Now gradually increase output until 1:1 capture returns – this is the stimulation threshold
- This is the lowest amount of energy that will reliably capture the heart and pace
- Set mA 1.5-3 times above this threshold value for safety
- Repeat for other circuit (A or V)
- Note: the stimulation threshold should be checked DAILY by the MD/NP when pacing is in use; in the event the threshold value is rising, discuss with surgeon to determine if alternative pacing system is needed
Determining the Sensing Threshold
- Note: the patient must have an underlying rhythm and tolerate a brief period without pacing to perform
- Set RATE at least 10 bpm below patient’s intrinsic heart rate (to allow sensing)
- Set OUTPUT at 0.1 mA (to avoid competitive pacing when device is made asynchronous briefly)
- Highlight SENSITIVITY (under Menu 1)
- Decrease SENSITIVITY: Slowly turn dial counter-clockwise until pace indicator flashes regularly (asynchronously)
- Increase SENSITIVITY: Slowly turn dial clockwise until sense indicator flashes again (when sensing resumes, this is the sensing threshold – the smallest complex that is reliably recognized, thus inhibiting the device)
- Set SENSITIVITY < half this threshold value for safety
- Repeat for other circuit (A or V)
- Restore previous rate and output values
Rapid Atrial (Overdrive) Pacing – PERFORMED BY MD/NP/CNS ONLY
- Used to “overdrive pace” and terminate some atrial reentrant tachyarrhythmias, especially atrial flutter
- ALWAYS VERIFY correct wire connections to avoid stimulating the ventricles at a rapid rate!
- Set maximal atrial output to promote capture (20 mA)
- Go to RAP menu (press MENU button three times)
- Dial in RAP rate 10-40 bpm above intrinsic atrial rate (default = 320)
- From RAP menu: press and hold SELECT button
- Rapid pacing will be delivered via atrial circuit
- Assess monitor for atrial “capture” at rapid rate
- Abruptly discontinue pacing by releasing SELECT button
- Examine rhythm and repeat if unsuccessful, consider: increasing rate and/or duration of stimulus, or gradually reducing pacing rate while holding SELECT